There are three fundamental types of risks associated with the uncertainty approach through making conformity or compliance decisions for tests which are based on meeting specification interval or regulatory limits.  Conformity decision rules can then be applied accordingly.

In summary, they are:

  1. Risk of false acceptance of a test result
  2. Risk of false rejection of a test result
  3. Shared risk

The basis of the decision rule is to determine an “Acceptance zone” and a “Rejection zone”, such that if the measurement result lies in the acceptance zone, the product is declared compliant, and, if it is in the rejection zone, it is declared non-compliant.  Hence, a decision rule documents the method of determining the location of acceptance and rejection zones, ideally including the minimum acceptable level of the probability that the value of the targeted analyte lies within the specification limits.

A straight forward decision rule that is widely used today is in a situation where a measurement implies non-compliance with an upper or lower specification limit if the measured value exceeds the limit by its expanded uncertainty, U

By adopting this approach, it should be emphasized that it is based on an assumption that the uncertainty of measurement is represented by a normal or Gaussian probability distribution function (PDF), which is consistent with the typical measurement results (being assumed the applicability of the Central Limit Theorem),

Current practices

When performing a measurement and subsequently making a statement of conformity, for example, in or out-of-specification to manufacturer’s specifications or Pass/Fail to a particular requirement, there can be only two possible outcomes:

  • The result is reported as conforming with the specification
  • The result is reported as not conforming with the specification

Currently, the decision rule is often based on direct comparison of measurement value with the specification or regulatory limits.  So, when the test result is found to be exactly on the dot of the specification, we would gladly state its conformity with the specification. The reason can be that these limits are deemed to have taken into account the measurement uncertainty (which is not normally true) or it has been assumed that the laboratory’s measurement value has zero uncertainty!  But, by realizing the fact that there is always a presence of uncertainty in all measurements, we are actually taking a 50% risk to have the actual or true value of the test parameter found outside the specification.  Do we really want to undertake such a high risky reporting? If not, how are we going to minimize our exposed risk in making such statement?